Design and construction of an optimized transmit/receive hybrid birdcage resonator to improve full body images of medium-sized animals in 7T scanner

Nibardo Lopez Rios, Philippe Pouliot, Konstantinos Papoutsis, Alexandru Foias, Nikola Stikov, Frédéric Lesage, Mathieu Dehaes et Julien Cohen-Adad

Article de revue (2018)

Document en libre accès dans PolyPublie et chez l'éditeur officiel
[img]
Affichage préliminaire
Libre accès au plein texte de ce document
Version officielle de l'éditeur
Conditions d'utilisation: Creative Commons: Attribution (CC BY)
Télécharger (1MB)
Afficher le résumé
Cacher le résumé

Abstract

The purpose of this work was to develop an optimized transmit/receive birdcage coil to extend the possibilities of a 7T preclinical MRI system to conduct improved full body imaging in medium-sized animals, such as large New Zealand rabbits. The coil was designed by combining calculation and electromagnetic simulation tools. The construction was based on precise mechanical design and careful building practice. A 16-leg, 20 cm long, 16 cm inner diameter, shielded quadrature hybrid structure was selected. Coil parameters were assessed on the bench and images were acquired on phantoms and rabbits. The results were compared to simulations and data obtained with an available commercial coil. An inexpensive assembly with an increase of 2 cm in useful inner diameter and 50 Omega matching with larger animals was achieved. A reduction in radiofrequency (RF) power demand of 31.8%, an improvement in image uniformity of 18.5 percentage points and an increase in signal-to-noise ratio of up to 42.2% were revealed by phantom image acquisitions, which was confirmed by in vivo studies. In conclusion, the proposed coil extended the possibilities of a preclinical 7T system as it improved image studies in relatively large animals by reducing the RF power demand, and increasing image uniformity and signal-to-noise ratio. Shorter scans and time under anesthesia or reduced RF exposure, resulting in better images and lower animal health risk during in vivo experiments, were achieved.

Mots clés

Animals; *Equipment Design; Magnetic Resonance Imaging/*instrumentation; Phantoms, Imaging; Rabbits

Sujet(s): 1900 Génie biomédical > 1900 Génie biomédical
1900 Génie biomédical > 1901 Technologie biomédicale
Département: Département de génie électrique
Organismes subventionnaires: Canada Research Chair in Quantitative Magnetic Resonance Imaging (JCA), Canadian Institute of Health Research, Fonds de Recherche du Québec - Santé, Fonds de Recherche du Québec - Nature et Technologies, Natural Sciences and Engineering Research Council of Canada, Quebec BioImaging Network
Numéro de subvention: CIHR FDN-143263, 28826, 32600, 2015-PR-182754, 435897-2013, 2015- 04672, 5886
URL de PolyPublie: https://publications.polymtl.ca/5008/
Titre de la revue: PLOS One (vol. 13, no 2)
Maison d'édition: PLOS
DOI: 10.1371/journal.pone.0192035
URL officielle: https://doi.org/10.1371/journal.pone.0192035
Date du dépôt: 17 déc. 2021 14:06
Dernière modification: 28 sept. 2024 17:17
Citer en APA 7: Lopez Rios, N., Pouliot, P., Papoutsis, K., Foias, A., Stikov, N., Lesage, F., Dehaes, M., & Cohen-Adad, J. (2018). Design and construction of an optimized transmit/receive hybrid birdcage resonator to improve full body images of medium-sized animals in 7T scanner. PLOS One, 13(2), e0192035 (14 pages). https://doi.org/10.1371/journal.pone.0192035