Benjamin Gervais, Aurelian Vadean, Maxime Raison
and Myriam Brochu
Article (2016)
|
Open Access to the full text of this document Published Version Terms of Use: Creative Commons Attribution Non-commercial No Derivatives Download (779kB) |
Abstract
This study presents a failure analysis of a femoral orthopedic implant. This implant is a locking compression plate that was fixed to a broken femur using two types of screws: locking and compression screws. The study elucidates the causes of an in situ premature failure of the plate and screws. Chemical analysis, hardness tests, and microstructural analysis confirmed that the implant was manufactured from cold-worked stainless steel 316L. The macro and micro fractographic analyses revealed that the failure mechanism was high-cycle fatigue and that the implant underwent approximately 10⁶ loading cycles before failure. A finite element analysis of the assembly indicated that the crack initiation sites are located in the region where the highest stresses are observed. This numerical analysis confirmed that walking induces the loading condition needed for this specific failure. According to this loading condition and to the material properties, the stress amplitude that initiated and propagated the crack is estimated to be in the range of 400 MPa. Several considerations, both mechanical and medical, are discussed in order to explain the failure and to improve the system durability. From an engineering perspective, implant geometry and installation procedure could be optimized in order to reduce the stress concentrations that developed near the crack origin.
Uncontrolled Keywords
| Department: | Department of Mechanical Engineering |
|---|---|
| Funders: | Fonds de Recherche du Québec-Nature et Technologies (FRQNT), Fondation de Polytechnique Montréal, Unité de participation et d’initiation à la recherche (UPIR) de Polytechnique Montréal |
| PolyPublie URL: | https://publications.polymtl.ca/4881/ |
| Journal Title: | Case Studies in Engineering Failure Analysis (vol. 5-6) |
| Publisher: | Elsevier |
| DOI: | 10.1016/j.csefa.2015.12.001 |
| Official URL: | https://doi.org/10.1016/j.csefa.2015.12.001 |
| Date Deposited: | 11 Aug 2022 13:56 |
| Last Modified: | 08 Jan 2026 23:50 |
| Cite in APA 7: | Gervais, B., Vadean, A., Raison, M., & Brochu, M. (2016). Failure analysis of a 316L stainless steel femoral orthopedic implant. Case Studies in Engineering Failure Analysis, 5-6, 30-38. https://doi.org/10.1016/j.csefa.2015.12.001 |
|---|---|
Statistics
Total downloads
Downloads per month in the last year
Origin of downloads
Dimensions