Abstract

This paper reports on a geometrically optimized implantable three-electrode Clark microsensor for measuring dissolved oxygen concentration. The sensor is based on a conventional concentric three electrode structure. In order to maximize the current between the working and counter electrode, sensor structures with different reference electrode sizes and shapes are simulated. Results show that the highest current is achieved for a half crescent electrode with a surface area half of the conventional full crescent reference electrode. The optimized microsensor structure is fabricated using silicon-based microtechnology. Electrochemical measurements performed with the sensor show that the sensor responds to different concentration of dissolved oxygen (DO2). A measured current of 0.87 µA between working and counter electrodes is reported for 96% of DO2 concentrations. The sensor shows a sensitivity of −0.0094 μA/%DO2 and an accuracy of 1.98% DO2 in measurement. A measurable output current at 2% DO2 of the microsensor demonstrates that the microsensor can have potential for low DO2 concentration measurement.

Mots clés

Département: Département de génie chimique
ISBN: 9781728144603
URL de PolyPublie: https://publications.polymtl.ca/82986/
Nom de la conférence: 2020 IEEE International Instrumentation and Measurement Technology Conference (I2MTC)
Lieu de la conférence: Dubrovnik, Croatia
Date(s) de la conférence: 2020-05-25 - 2020-05-28
DOI: 10.1109/i2mtc43012.2020.9129243
URL officielle: https://doi.org/10.1109/i2mtc43012.2020.9129243
Date du dépôt: 01 oct. 2026 14:20
Dernière modification: 01 oct. 2026 14:20
Citer en APA 7: Nosrati, M., Vieira, D., Joseph Harvey, E., Merle, G., & Bhadra, S. (mai 2020). Development of a Clark microsensor for low concentration dissolved oxygen monitoring [Communication écrite]. 2020 IEEE International Instrumentation and Measurement Technology Conference (I2MTC), Dubrovnik, Croatia (5 pages). https://doi.org/10.1109/i2mtc43012.2020.9129243

Statistiques

Dimensions

Actions réservées au personnel

Afficher document Afficher document