S. Vallières, Charles Bienvenue, P. Puyuelo-Valdes, M. Salvadori, E. d’Humières, F. Schiettekatte et P. Antici
Article de revue (2019)
Document publié alors que les auteurs ou autrices n'étaient pas affiliés à Polytechnique Montréal
Un lien externe est disponible pour ce documentAbstract
In this work, we calibrate the newly developed EBT-XD radiochromic films (RCFs) manufactured by Gafchromicᵀᴹ using protons in the energy range of 4–10 MeV. Irradiation was performed on the 2 × 6 MV tandem linear accelerator located at the Université de Montréal. The RCFs were digitized using an Epson Perfection V700 flatbed scanner using both the red-green-blue and grayscale channels. The proton fluences were measured with Faraday cups calibrated in absolute terms. The linear energy transfer function within the active layer of the films was calculated using the mass stopping power tables coming from the PSTAR database from the National Institute of Standards and Technology (NIST) to allow retrieval of the deposited dose. We find that the calibration curves for 7 and 10 MeV protons are nearly equivalent. The 4 MeV calibration curves exhibit a quenching effect due to the Bragg peak that falls close to the active layer. A linearization of this energy dependence was developed using a semiempirical parametric model to allow the generation of calibration curves for any incident proton energy within the present range. Excellent correspondence (<5% dose difference for the same netOD) of the 10 MeV calibration curves was noted when compared to existing high-energy proton (148.2 MeV) calibration curves reported in the literature. Our calibration extends the range of operation of EBT-XD films to low-energy proton beam dosimetry.
Mots clés
| Département: | Non applicable |
|---|---|
| Organismes subventionnaires: | Fonds de Recherche du Québec - Nature et Technologie, NSERC / CRSNG - Discovery Grant, Alexander-Graham-Bell Graduate Scholarship, Compute Canada, Canada Foundation for Innovation (CFI), French National Research Agency (ANR), ANR-10-IDEX-03-02 |
| Numéro de subvention: | 174726, 2016-PR-189974, RGPIN-2018-05772, pve-323-ac, ANR-10-IDEX-03-02 |
| URL de PolyPublie: | https://publications.polymtl.ca/67512/ |
| Titre de la revue: | Review of Scientific Instruments (vol. 90, no 8) |
| Maison d'édition: | American Institute of Physics |
| DOI: | 10.1063/1.5109644 |
| URL officielle: | https://doi.org/10.1063/1.5109644 |
| Date du dépôt: | 13 août 2025 17:01 |
| Dernière modification: | 13 août 2025 17:01 |
| Citer en APA 7: | Vallières, S., Bienvenue, C., Puyuelo-Valdes, P., Salvadori, M., d’Humières, E., Schiettekatte, F., & Antici, P. (2019). Low-energy proton calibration and energy-dependence linearization of EBT-XD radiochromic films. Review of Scientific Instruments, 90(8), 083301 (8 pages). https://doi.org/10.1063/1.5109644 |
|---|---|
Statistiques
Dimensions